@article{2943a439cf614890ab65e090a17f20ee,
title = "Using Tracheal Pressure to Trigger the Ventilator and Control Airway Pressure During Continuous Positive Airway Pressure Decreases Work of Breathing",
author = "Gabriel Messinger and Banner, \{Michael J.\} and Blanch, \{Paul B.\} and Layon, \{Abraham Joseph\}",
note = "We evaluated the difference in work of breathing (WOB) during spontaneous ventilation with continuous positive airway pressure (CPAP) among three methods of triggering the ventilator: conventional pressure triggering, tracheal pressure triggering, and flow-by triggering. In an in vitro model of the respiratory system consisting of a bellows (lungs) in a plastic canister (chest wall), spontaneous ventilation was simulated with a piston-driven pump (respiratory muscles).",
year = "1995",
month = aug,
day = "1",
doi = "10.1378/CHEST.108.2.509",
language = "American English",
volume = "108",
journal = "Chest",
issn = "1931-3543",
publisher = "American College of Chest Physicians",
number = "2",
}